    Computed Radiography of Chest Radiographic Imaging: A Phantom Study of Optimal Imaging Parameters and Radiation Dose

    Access Status
    Fulltext not available
    Authors
    Sun, Zhonghua
    Siddiqu, S.
    Ng, K.
    Ramli, K.
    Davidson, R.
    Date
    2007
    Type
    Conference Paper
    
    Metadata
    Show full item record
    Citation
    Sun, Zhonghua and Siddiqu, Salim and Ng, Kwan-hoong and Ramli, Khadijah and Davidson, Robert. 2007. : Computed Radiography of Chest Radiographic Imaging: A Phantom Study of Optimal Imaging Parameters and Radiation Dose, in Sison, Lilian J. (ed), 5th SouthEast Asian Congress of Medical Physics, 21-23 Nov 2007. Manila: Philippine Organization of Medical Physicists.
    Source Title
    5th SouthEast Asian Congress of Medical Physics
    Source Conference
    5th SouthEast Asian Congress of Medical Physics
    Faculty
    Department of Medical Imaging Science
    Division of Engineering, Science and Computing
    Faculty of Science
    URI
    http://hdl.handle.net/20.500.11937/34847
    Collection
    • Curtin Research Publications
    Abstract

    Introduction: In chest radiographic imaging, computed radiography (CR) has been replacing the conventional film-screen imaging technology. Selection of the appropriate radiation quality is an important aspect of optimisation for chest imaging as it affects both image quality and patient dose. We aimed to investigate the optimal imaging parameters for chest radiographic imaging using CR system, with regards to the relationship between tube potential and radiation dose.Methods: The study was performed on a chest phantom made from sheets of plastic tubing, copper and aluminium, which were shaped to resemble frontal radiographic projections of human thoracic structures. Regional test objects were incorporated into the chest phantom for image quality assessment in the lungs, heart and retrodiaphragmatic areas. Chest phantom images were taken using CR on an Agfa system. Imaging parameters were selected with mAs ranging from 0.5, 1.0, 2.0, 4.0 and 8.0, and tube potential ranging from 100, 110 and 120 kV. Entrance skin dose (mGy) was measured using a solid state detector. Quantitative measurements of image quality were performed at 7 regions of interest to determine the relationship between image noise and imaging parameters.Results: The radiation dose increased significantly when the mAs was increased (p<0.05). However, no significant difference of radiation dose was found among the varying kV values (p>0.05). The image noise measured in different regions of interest increased significantly (p<0.001) when the kV was increased, with 120 kV generating the lowest noise value. Subjective assessment of image quality showed that chest radiographic images acquired with 2 mAs and 4 mAs were acceptable for diagnostic purpose.Discussion: Optimisation of CR imaging means that image quality must satisfy the diagnostic requirements for making a correct diagnosis at the lowest possible radiation dose. The findings in this study suggested that image quality improved with increased kV without any significant increase in radiation dose.Conclusion: Our results showed that CR chest radiography produced optimum image quality following the protocol of 120 kV and 2 mAs. The chest image quality in CR imaging can be improved by increasing the tube potential while maintaining relatively low dose.

    Related items

    Showing items related by title, author, creator and subject.

    • Investigation of optimal parameters for chest radiograph imaging using film-screen and computed radiography
      Sun, Zhonghua; Siddiqu, Salim; Ng, K.; Ramli, K.; Davidson, R. (2007)
      Objective: Digital imaging is gradually replacing the conventional film-screen (FS) imaging technology. This is mainly due to the perceived advantages of digital image processing, electronic archiving, teleradiology and ...
    • Coronary CT angiography : radiation dose measurements and image quality assessments
      Sabarudin, Akmal (2012)
      Prospective ECG-triggering is regarded as one of the most effective approaches for reduction of radiation dose to patients during coronary CT angiography (CCTA). This study was conducted to investigate the diagnostic ...
    • Chest computed radiography imaging parameters for nodule detection: a comparison of image quality and entrance skin dose
      Sun, Zhonghua (2014)
      This study was performed on an anthropomorphic phantom to investigate the performance of three computed radiography (CR) systems with regard to the detection of simulated lung nodules and entrance skin dose. Forty-five ...
    Advanced search

    Browse

    Communities & CollectionsIssue DateAuthorTitleSubjectDocument TypeThis CollectionIssue DateAuthorTitleSubjectDocument Type

    My Account

    Admin

    Statistics

    Most Popular ItemsStatistics by CountryMost Popular Authors

    Follow Curtin

    • 
    • 
    • 
    • 
    • 

    CRICOS Provider Code: 00301JABN: 99 143 842 569TEQSA: PRV12158

    Copyright | Disclaimer | Privacy statement | Accessibility

    Curtin would like to pay respect to the Aboriginal and Torres Strait Islander members of our community by acknowledging the traditional owners of the land on which the Perth campus is located, the Whadjuk people of the Nyungar Nation; and on our Kalgoorlie campus, the Wongutha people of the North-Eastern Goldfields.