dc.identifier.citationMcCann, Kirstie and Lim, Victor and Tomlinson, Guy and Way, Jason and Alshamakhi, Ahmed and Ong, Jeanette and Sun, Zhonghua. 2007. Optimisation of radiation dose and image quality in 3D multi-detector CT angiography in abdominal aortic aneurysm: in vitro aorta phantom study. The Radiographer 54 (1): 14-18.
dc.identifier.urihttp://hdl.handle.net/20.500.11937/47839
dc.description.abstract

We aimed to investigate relationship of radiation dose and image quality in 3D multi-detector CT angiography of abdominal aortic aneurysm based on an aorta phantom. The study was conducted using a human aorta phantom modelled on a patient with an abdominal aortic aneurysm. To perform the scans of the aorta phantom it was put in a specially designed, sealable perspex container, which was filled with iodinated contrast medium. A series of scans were performed using different protocols to test the effect on dose and image quality. These were: the variable detectors (4-, 32- and 64-slice) used in image acquisition, mA and pitch values. A line profile analysis was conducted at three separate locations for each different protocol, namely, the renal artery, superior mesenteric artery, and abdominal aortic aneurysm to determine the degree of image artefacts using standard deviation, which was measured on the virtual intravascular endoscopy images. The effective dose was also calculated for each scanning protocol. Our results indicate that the scans performed on 64 detectors multi-detector CT produce better image quality compared with 32 and 4 detectors. Increasing scanning parameters such as mA enhances image quality but it increases patients? radiation dose too. It is also noted that the pitch is inversely proportional to radiation dose, resulting in degradation of image quality. From our results, a 64 multi-detector CT scanner with a scanning protocol of 120kVp, 70mA, and a pitch of 0.641 is ideal for imaging of abdominal aortic aneurysms because it allows for fewer artefacts and less radiation dose when compared to other scanning protocols.

dc.publisherAustralian Institute of Radiography
dc.relation.urihttp://search.informit.com.au/fullText;dn=87088522;res=MEDITEXT
dc.subjectimage quality
dc.subjectMulti-detector scan
dc.subjectradiation dose
dc.subjectartefacts
dc.subjectaneurysm
dc.titleOptimisation of radiation dose and image quality in 3D multi-detector CT angiography in abdominal aortic aneurysm: in vitro aorta phantom study
dc.typeJournal Article
dcterms.source.volume54
dcterms.source.number1
dcterms.source.monthapr
dcterms.source.startPage14
dcterms.source.endPage18
dcterms.source.titleThe Radiographer
curtin.note

The Australian Institute of Radiography holds the copyright on papers published in The Radiographer. The Australian Institute of Radiography website is available at http://www.air.asn.au/

curtin.note

© The Australian Institute of Radiography The Australian Institute of Radiography website is available at http://www.air.asn.au/

curtin.note

Access to the journal The Radiographer is available from www.minniscomms.com.au/radiographer/

curtin.departmentDepartment of Imaging and Applied Physics
curtin.identifierEPR-2831
curtin.accessStatusOpen access
curtin.facultyDepartment of Medical Imaging Science
curtin.facultyDivision of Engineering, Science and Computing
curtin.facultyFaculty of Science


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record