Portuguese Study Of Mean Glandular Dose In Mammography And Comparison With European References
Access Status
Authors
Date
2018Type
Metadata
Show full item recordCitation
Source Title
School
Collection
Abstract
To characterise the mean glandular dose (MGD) in a sample of healthcare providers for digital mammography in Portugal. To compare the achieved values with European references. The MGD was measured on a poly-methyl-methacrylate phantom (45 mm) for each system using dosimeters. In addition, MGD was estimated using exposure settings collected from mammography exams in clinical context. Data were collected from 25 computed-radiography systems (CR) and 13 integrated digital (DR). For both measurements (phantom and clinical exposures), the average MGD for CR was higher compared to the DR. For CR the mean MGD was 1.85 mGy (CC projection) and 2.10 mGy (MLO projection). For DR systems the corresponding values were 1.54 mGy (CC) and 1.68 mGy (MLO). The average MGD obtained using both methods and for both technologies is within the acceptable reference range proposed by European guidelines (<2.5 mGy). Dose Reference Levels implementation should be the next step to optimise mammography practice in Portugal.
Related items
Showing items related by title, author, creator and subject.
-
Fartaria, M.; Reis, Claudia; Pereira, J.; Pereira, M.; Cardoso, J.; Santos, L.; Oliveira, C.; Holovey, V.; Pascoal, A.; Alves, J. (2016)The aim of this paper is the characterization of four thermoluminescence detectors (TLD), namely, LiF:Mg,Ti, LiF:Mg,Cu,P, Li2B4O7:Mn and Li2B4O7:Cu for the measurement of the entrance surface air kerma (ESAK) and estimation ...
-
Metsälä, E.; Meystre, N.; Jorge, J.; Henner, A.; Kukkes, T.; Sá dos Reis, Claudia (2017)Objectives: This study aims to identify European radiographers’ challenges in clinical performance in mammography and the main areas of mammography that require more and better training. Methods: An extensive search was ...
-
Costa, S.; Oliveira, E.; Reis, Claudia; Viegas, S.; Serranheira, F. (2014)Objectives: Identify radiographers’ postures during frequent mammography procedures related to the mammography equipment and patient characteristics. Methods: A postural task analysis was performed using images acquired ...