dc.identifier.citationGill, M. and Vijayananthan, A. and Kumar, G. and Jayarani, K. and Ng, K. and Sun, Z. 2015. Use of 100 kV versus 120 kV in computed tomography pulmonary angiography in the detection of pulmonary embolism: effect on radiation dose and image quality. Quantitative imaging in medicine and surgery. 5 (4): pp. 524-533.
dc.identifier.urihttp://hdl.handle.net/20.500.11937/28934
dc.identifier.doi10.3978/j.issn.2223-4292.2015.04.04
dc.description.abstract

Objective: To determine the effective radiation dose and image quality resulting from 100 versus 120 kilovoltage (kV) protocols among patients referred for computed tomography pulmonary angiography (CTPA). Methods: Sixty-six patients with clinical suspicion of pulmonary embolism (PE) were prospectively enrolled. Two CTPA protocols (group A: n=33, 100 kV/115 mAs; group B: n=33, 120 kV/90 mAs) were compared. Two experienced radiologists assessed image quality in terms of diagnostic performance and effect of artefacts. Image quality parameters [CT attenuation, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR)] and effective radiation dose between the two protocols were compared. Results: The contrast enhancement in central and peripheral pulmonary arteries was significantly higher in group A than in group B (P<0.001) with the identical SNR (P=0.26), whereas the CNR was significantly higher in group A than in group B (P<0.001). The effective radiation dose for the 100 and 120 kV scans was 3.2 and 6.8 mSv, respectively. Conclusions: Reducing the tube voltage from 120 to 100 kV in CTPA allows a significant reduction of radiation dose without significant loss of diagnostic image quality.

dc.titleUse of 100 kV versus 120 kV in computed tomography pulmonary angiography in the detection of pulmonary embolism: effect on radiation dose and image quality.
dc.typeJournal Article
dcterms.source.volume5
dcterms.source.number4
dcterms.source.startPage524
dcterms.source.endPage533
dcterms.source.issn2223-4292
dcterms.source.titleQuant Imaging Med Surg