Optimum Tube Voltage for Pelvic Direct Radiography: A Phantom Study
Access Status
Authors
Date
2015Type
Metadata
Show full item recordCitation
Source Title
ISSN
School
Remarks
This open access article is distributed under the Creative Commons license http://creativecommons.org/licenses/by-nc-nd/2.5/ZA
Collection
Abstract
Pelvic radiography is a frequently performed radiological examination. Its average effective dose (E) is 0.53 mSv which is comparable to the annual per caput dose from diagnostic radiology, 0.6 mSv. However, existing studies on optimum tube potential for pelvic X-rays tend to be limited to screen-film and computed radiography. The purpose of this study was to determine the tube voltage for dose-image optimisation in pelvic direct radiography (DR). Fifty-four pelvic phantom images were acquired using 50-135 kV at 5 kV increments (three images taken at each kV level) and milliampere seconds determined by automatic exposure control. The signal-to-noise ratio (SNR) and dose were measured for each image. Figure of merit (FOM) defined as the ratio of SNR2 to E was used to determine the optimum tube potential. The FOM indicates 135 kV is the optimum setting for pelvic DR. Using the European Commission tube voltage recommendation (75-90 kV) as a reference point, there was only a slight (5.56%) decrease of image quality in the femoral neck region at 135 kV. However, its E was 0.054 mSv. This appreciable dose reduction potential could be attributed to the improvement of detective quantum efficiency and image processing technology of the recent DR system.
Related items
Showing items related by title, author, creator and subject.
-
Sun, Zhonghua; Siddiqu, S.; Ng, K.; Ramli, K.; Davidson, R. (2007)Introduction: In chest radiographic imaging, computed radiography (CR) has been replacing the conventional film-screen imaging technology. Selection of the appropriate radiation quality is an important aspect of optimisation ...
-
Sun, Zhonghua; Siddiqu, Salim; Ng, K.; Ramli, K.; Davidson, R. (2007)Objective: Digital imaging is gradually replacing the conventional film-screen (FS) imaging technology. This is mainly due to the perceived advantages of digital image processing, electronic archiving, teleradiology and ...
-
Sabarudin, Akmal; Sun, Zhonghua; Mustafa, Z.; Nassir, K.; Hamid, H. (2014)This phantom study was designed to compare the radiation dose in thoracic and abdomen-pelvic CT scans with and without use of tube current modulation (TCM). Effective dose (ED) and size-specific dose estimation (SSDE) ...