    Optimum Tube Voltage for Pelvic Direct Radiography: A Phantom Study

    236123_236123.pdf (1.491Mb)
    Access Status
    Open access
    Authors
    Jacobs, S.
    Kuhl, L.
    Xu, G.
    Powell, R.
    Paterson, D.
    Ng, Curtise Kin Cheung
    Date
    2015
    Type
    Journal Article
    
    Metadata
    Show full item record
    Citation
    Jacobs, S. and Kuhl, L. and Xu, G. and Powell, R. and Paterson, D. and Ng, C.K.C. 2015. Optimum Tube Voltage for Pelvic Direct Radiography: A Phantom Study. The South African Radiographer. 53 (2): pp. 15-19.
    Source Title
    The South African Radiographer
    ISSN
    0258-0241
    School
    Department of Medical Radiation Sciences
    Remarks

    This open access article is distributed under the Creative Commons license http://creativecommons.org/licenses/by-nc-nd/2.5/ZA

    URI
    http://hdl.handle.net/20.500.11937/8929
    Collection
    • Curtin Research Publications
    Abstract

    Pelvic radiography is a frequently performed radiological examination. Its average effective dose (E) is 0.53 mSv which is comparable to the annual per caput dose from diagnostic radiology, 0.6 mSv. However, existing studies on optimum tube potential for pelvic X-rays tend to be limited to screen-film and computed radiography. The purpose of this study was to determine the tube voltage for dose-image optimisation in pelvic direct radiography (DR). Fifty-four pelvic phantom images were acquired using 50-135 kV at 5 kV increments (three images taken at each kV level) and milliampere seconds determined by automatic exposure control. The signal-to-noise ratio (SNR) and dose were measured for each image. Figure of merit (FOM) defined as the ratio of SNR2 to E was used to determine the optimum tube potential. The FOM indicates 135 kV is the optimum setting for pelvic DR. Using the European Commission tube voltage recommendation (75-90 kV) as a reference point, there was only a slight (5.56%) decrease of image quality in the femoral neck region at 135 kV. However, its E was 0.054 mSv. This appreciable dose reduction potential could be attributed to the improvement of detective quantum efficiency and image processing technology of the recent DR system.

    Related items

    Showing items related by title, author, creator and subject.

    • Computed Radiography of Chest Radiographic Imaging: A Phantom Study of Optimal Imaging Parameters and Radiation Dose
      Sun, Zhonghua; Siddiqu, S.; Ng, K.; Ramli, K.; Davidson, R. (2007)
      Introduction: In chest radiographic imaging, computed radiography (CR) has been replacing the conventional film-screen imaging technology. Selection of the appropriate radiation quality is an important aspect of optimisation ...
    • Investigation of optimal parameters for chest radiograph imaging using film-screen and computed radiography
      Sun, Zhonghua; Siddiqu, Salim; Ng, K.; Ramli, K.; Davidson, R. (2007)
      Objective: Digital imaging is gradually replacing the conventional film-screen (FS) imaging technology. This is mainly due to the perceived advantages of digital image processing, electronic archiving, teleradiology and ...
    • Radiation dose reduction in thoracic and abdomen-pelvic CT using tube current modulation: A phantom study
      Sabarudin, Akmal; Sun, Zhonghua; Mustafa, Z.; Nassir, K.; Hamid, H. (2014)
      This phantom study was designed to compare the radiation dose in thoracic and abdomen-pelvic CT scans with and without use of tube current modulation (TCM). Effective dose (ED) and size-specific dose estimation (SSDE) ...
    Advanced search

    Browse

    Communities & CollectionsIssue DateAuthorTitleSubjectDocument TypeThis CollectionIssue DateAuthorTitleSubjectDocument Type

    My Account

    Admin

    Statistics

    Most Popular ItemsStatistics by CountryMost Popular Authors

    Follow Curtin

    • 
    • 
    • 
    • 
    • 

    CRICOS Provider Code: 00301JABN: 99 143 842 569TEQSA: PRV12158

    Copyright | Disclaimer | Privacy statement | Accessibility

    Curtin would like to pay respect to the Aboriginal and Torres Strait Islander members of our community by acknowledging the traditional owners of the land on which the Perth campus is located, the Whadjuk people of the Nyungar Nation; and on our Kalgoorlie campus, the Wongutha people of the North-Eastern Goldfields.